3-104 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Errors and Failures Dialog Box
0xBFFC0896 ERR_ERROR_READING_FILE
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"Error reading data
from a file"
0xBFFC0897 ERR_ERROR_WRITING_FILE
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"Error writing data to
a file"
Error Queue Errors
0xBFFC08A0 ERR_ERROR_QUEUE_OVERFLOW
This occurs when the error queue reaches its
maximum size limit which is currently defined as
30 errors. Reading the error queue cleared this
error and empties the error queue to allow errors
to be properly queued.
"Error queue error"
0xBFFC08A1 ERR_PAGE_LOCK_DEADLOCK
This may occur if two applications are exercising a
single E6432 and accessing common data. Make
sure only one application is running that
interfaces with the E6432.
"Page register lock is
deadlocked"
Frequency and Power Setting Errors (continued)
0xBFFC08B0 ERR_TIMEOUT_WAITING_FOR_ABORT
This error should not be generated under normal
operation. However, if the user is using more than
one application to control the E6432 at the same
time, this type of error could be generated. Verify
that there is only one application interfacing with
the unit at any given time. Contact an Agilent
Technologies Service Center for more information.
"Timeout Error While
Waiting For List to
Abort"
0xBFFC08B1 ERR_WAITING_FOR_SETTLED
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"Error While Waiting
For Settled Interrupt"