User`s guide

3-102 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Errors and Failures Dialog Box
0xBFFC088E ERR_ALC_HIGH_UNLEVELED_TRANSIENT
This error occurs if power is set above the
maximum specified power level when using
internal leveling. Check to ensure the proper
power level is being requested.
If using external leveling, this could occur if the
power is set above any of the internal maximum
power limits, which could be caused by the
external hardware configuration and power levels
being selected in association with them. Check the
external configuration and ensure that the
returned power level within the loop and that
being selected are within the proper operational
range of the E6432.
"Transient ALC High
Power Unleveled
Error"
0xBFFC088F ERR_ALC_LOW_UNLEVELED_TRANSIENT
This error indicates that the unleveled interrupt is
transient and that it isn’t always set. This could
occur under the following conditions when either
the provided return signal is oscillating or the
selected level is on the edge of the system
performance, causing the leveled interrupt to
oscillate between being set and not-set.
If power is set beyond the minimum specified
power level when using internal leveling. Check to
ensure the proper power level is being requested.
If using external leveling this could occur if the
power is set beyond any of the internal minimum
power limits, which could be caused by the
external hardware configuration and power levels
being selected in association with them. Check the
external configuration and ensure that the
returned power level within the loop and that
being selected are within the proper operational
range of the E6432.
This error could also be caused by external
modulation inputs that are not within the
allowable range. Verify that all external
modulation is within the specified limits for proper
operation.
"Transient ALC Low
Power Unleveled
Error"