User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 3-101
Soft Front Panel Help
Errors and Failures Dialog Box
0xBFFC0889 ERR_LIST_STILL_RUNNING
This error will occur when the user attempts to set
frequency and power while a list is being executed.
The user should either wait for the list to
complete, or abort the list prior to attempting to
set frequency and power.
"List is still running"
0xBFFC088A ERR_EXTERNAL_REF100_UNLOCKED_TRANS
IENT
This error indicates that the E6432 is not always
able to lock with the external reference source.
The error is transient indicating it isn’t always set.
Check the external reference connection for a
possible loose connection and verify that the
external reference source is constantly putting out
the proper signal.
"Transient External
10 MHz Reference
Unlock Error"
0xBFFC088C ERR_UWPLL_UNLOCKED_TRANSIENT
For some conditions, when an external reference is
used and the reference source is reapplied (for
example, powered down then back up, or
disconnected then reconnected) a transient may
occur. Reading the error queue should clear this
error.
If the reference is constantly applied and this error
is not cleared when the error is read, this may
indicate a hardware failure. If this error occurs
when an internal reference is selected, this may
also indicate a hardware failure.
"Transient Microwave
Phase Locked Loop
Unlock Error"
0xBFFC088D ERR_PAREN_UNLOCKED_TRANSIENT
This error could occur, for some frequency value
settings, when the system is driven in FM with
extreme modulated power. Ensure that any
modulated input is within the specified
operational range for the system.
"Transient Paren
Unlock Error"