User`s guide

3-100 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Errors and Failures Dialog Box
0xBFFC0885 ERR_EXTERNAL_REF100_UNLOCKED
This error indicates that the E6432 is unable to
lock with the external reference source. Check the
external reference connection and also verify that
the external reference source is putting out the
proper signal.
"External 100 MHz
Reference Unlock
Error"
0xBFFC0886 ERR_UNLOCK_IN_LIST
This error could occur while running a list, for
some frequency value settings, when the system is
driven in FM with extreme modulated power.
Ensure that any modulated input is within the
specified operational range for the system.
"Unleveled in List
Error"
0xBFFC0887 ERR_UNLEVEL_IN_LIST
This error could occur if using internal leveling
and the power is set beyond the minimum or
maximum specified power limits within a list
point. Check to ensure the proper power level is
being requested for all the list points.
This error could occur if using external leveling
and the power is set beyond any of the internal
minimum or maximum power limits; this could be
caused by the external hardware configuration or
the power levels being selected in association with
them. Check the external configuration and
ensure that the returned power level within the
loop and that being selected are within the proper
operational range of the E6432.
"Unlevel in List Error"
0xBFFC0888 ERR_TIMEOUT_WAITING_FOR_SETTLED
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"Timeout Error While
Waiting For Settled
Interrupt”