User`s guide

Agilent E6432A Microwave Synthesizer Users Guide 3-95
Soft Front Panel Help
Errors and Failures Dialog Box
0xBFFC0B04 ERR_OPEN_A24_FAILED
The address of the E6432 specified in the
HPE6432_init call is probably incorrect. If the
address specified is correct, this could also be a
Windows resource error. Monitor Windows
resources to verify that there is enough space to
allocate memory for this process.
"VISA Open A24
Failed (see next error)"
0xBFFC0B05 ERR_SET_TIMEOUT_A24_FAILED
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"VISA Set A24
Timeout Failed (see
next error)"
0xBFFC0B06 ERR_MAP_ADDRESS_A24_FAILED
This could be a Windows resources error. Monitor
the Windows resources to verify that there is
enough space to allocate memory for this process.
"VISA Map A24
Address Failed (see
next error)"
0xBFFC0907 ERR_INSTALL_HANDLER_FAILED
The interface for this system is not one of the
supportable interfaces, such as GPIB which
doesn’t support interrupt handling. Verify that a
supportable interface is being used.
"VISA Install Handler
Failed (see next error)"
0xBFFC0908 ERR_ENABLE_EVENT_FAILED
The interface for this system is not one of the
supportable interfaces, such as GPIB which
doesn’t support interrupt handling. Verify that a
supportable interface is being used.
"VISA Enable Event
Failed (see next error)"