User`s guide
3-92 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Errors and Failures Dialog Box
Read and Clear Error Queue
Read and Clear Error Queue, available from the Errors and Failures
Dialog Box, queries the synthesizer and reads all of the current error
messages from the synthesizer’s error queue, displays the errors in the
display buffer, and clears the synthesizer’s error queue. If error
messages are currently displayed in the Errors and Failures Dialog
Box, new messages are appended to the end of the list being displayed.
Reading error messages that begin with ERR from the error queue
remove them permanently. Reading error messages that begin with
FAIL remain on the error queue and are only removed when the
system’s power is cycled.
Messages that begin with FAIL are considered hardware failures and
the synthesizer must be repaired. All other error messages can be
resolved by the user. For hints on resolving an error message, refer to
“Error-Code and Fail-Code Messages” on page 3-93.
When the power to the synthesizer is cycled off and on, all errors in the
synthesizer’s error queue are lost.
TIP Because error messages are appended to the display each time the Read
and Clear Error Queue button is selected, the Clear Display button
should be selected to remove old error messages; a list of error messages
that are no longer valid may be displayed if this is not done.
Related Topics
Pull Down View Menu
Soft Front Panel Help
Errors and Failures Dialog Box
Error-Code and Fail-Code Messages
To Display a List of the Synthesizer’s Error Queue Messages
To Print a List of the Synthesizer’s Error Queue Messages
HPE6432_error_query