User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 3-81
Soft Front Panel Help
List Point Calculator Dialog Box
Related Topics
Soft Front Panel Help
List Point Calculator Dialog Box
Don’t Specify the Start, Stop, Step, or the # of Steps
Parameter
Don’t Specify, available from the List Point Calculator Dialog Box,
designates that the Start, Stop, Step, or the # of Steps parameter is not
specified. In this way, the number of list points that are created for a
given frequency range or power range can be calculated using only
three of the four parameters; one of the four parameters is always
disabled.
• Don’t Specify Step
• Don’t Specify Start
• Don’t Specify Stop
• Don’t Specify # of Steps
Don’t Specify Step
Selecting Don’t Specify Step, on the List Point Calculator Dialog Box,
designates that the Step parameter is not specified when calculating a
frequency and ALC power list.
For example, produce a list with ten steps over a frequency range of
100 MHz to 150 MHz with corresponding ALC Power levels from 0 dBm
to 10 dBm. The list that is produced has a frequency step size of 5 MHz
and a 1 dB ALC power change per step.