User`s guide

Agilent E6432A Microwave Synthesizer Users Guide 3-73
Soft Front Panel Help
List Dialog Box
Switch/blanking time is established by the following criteria:
350 us for all frequencies 560 MHz or less
350 us for all frequencies above 560 MHz with long switch/blanking
mode set (long blanking is typically used when external leveling is
enabled or when using low ALC bandwidth)
150 us for all frequencies above 560 MHz with normal
switch/blanking mode set
50 us for all frequencies above 560 MHz with power-only mode set
20 ms is added, to each of the times listed above, any time the step
attenuator (Option 1E1) is changed
Once switch/blanking time is completed, the settling time begins.
Settling time is user-definable and can be adjusted between a minimum
and maximum value; longer periods of settling time can be specified in
order to gain additional accuracy. Both switch/blanking time and
settling time are used every time a new frequency or power is set up in
the synthesizer. (For more information, refer to “Settling Time” on page
3-43.)
In summary, there is a delay time that is required by the synthesizer to
change between frequencies or power or both. This delay time is a
combination of the switch/blanking time and settling time. The
switch/blanking time is dependent on the criteria listed above while the
settling time is user-definable and dependent on the accuracy required
of the final signal. If RF blanking is on during switch/blanking time,
you do not see the effects on the RF output, but if RF blanking is off
during switch/blanking time, you see all of the effects on the signal that
might include spurious signals, harmonics, and other glitches. Whether
RF blanking is on or off has no effect on settling time, it only affects the
RF output during switch/blanking time.
Related Topics
Soft Front Panel Help
List Dialog Box Controls
Flags - of a List Point
Blanking - of a List Point
Settling Time
HPE6432_WriteListPoints
HPE6432_SetLongBlankingState