User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 3-63
Soft Front Panel Help
List Dialog Box
Related Topics
Soft Front Panel Help
List Dialog Box Controls
Trig In (0, 1, 2) - List Playing Control
Trig In #, available from the List Dialog Box, specifies the action that a
Trig In trigger has on the current list.
Trig In triggers can come from the Trig In connector on the hardware
front panel or one of the eight-shared VXI backplane TTL triggers
(TTLTRG0-TTLTRG7). Refer to Trigger Input, available from the
Configuration Dialog Box, to control which signal will be used as the
Trig In trigger.
Trig In 0 – Auto Trigger (Ignore)
If the Repeat check box, available from the List Dialog Box, is selected,
this setting runs the list repeatedly until stopped and does not require a
Trig In trigger. The list can be stopped using the Stop button available
from the List Dialog Box.
If the Repeat check box, available from the List Dialog Box, is not
selected, this setting runs the list once automatically and does not
require a Trig In trigger.
Trig In 1 – Wait for Trig In to Step to Next Point
This setting specifies that the synthesizer is to step to the next point in
the list each time a Trig In trigger is received.
Trig In 2 – Wait for Trig In to Run Full List
This setting specifies that the synthesizer runs through the entire list
automatically after receiving a single Trig In trigger.
TIP These same modes can be specified programmatically by setting the
featureBits parameter using the HPE6432_RunList() function.