3-54 Agilent E6432A Microwave Synthesizer Users Guide
Soft Front Panel Help
Configuration Dialog Box
This is an output trigger (that can also be used as a marker) and is
asserted during dwell time; the value of the dwell time controls how
long the trigger outputs (Sync Out trigger and Trig Out trigger) are
asserted. The Sync Out trigger is produced after each point in the list
has settled if the point has a Sync Out bit enabled. This output trigger
can be directed to the Sync Out connector on the hardware front panel
or one of the eight-shared VXI backplane TTL triggers
(TTLTRG0-TTLTRG7) or both at the same time.
When the Sync Out trigger is directed to the front panel connector, the
following apply:
Related Topics
Soft Front Panel Help
Configuration Dialog Box Controls
HPE6432_SetExtSyncOutput
Sync Out (VXI Backplane)
Sync Out (VXI Backplane), available from the Configuration Dialog
Box, provides a way to control which VXI backplane line will have the
Sync Out trigger.
FRONT_OUT_POS = 16 Front Panel + The Sync Out trigger is
generated with a positive
polarity.
FRONT_OUT_NEG = 48 Front Panel – The Sync Out trigger is
generated with a negative
polarity.
FRONT_OUT_OFF = 0 Off No Sync Out trigger is output.