User`s guide
3-50 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Configuration Dialog Box
Related Topics
Soft Front Panel Help
Configuration Dialog Box Controls
HPE6432_SetTriggerInput
Settling Time
Dwell Time
Trigger Out (Front Panel)
triggerInputSource
Source Description Source of Input Trigger
Trig In From the Trig In connector on
the hardware front panel of the
synthesizer.
FRONT_IN_POS = 24 Front Panel
Pos Edge
The trigger is asserted with a
positive edge polarity.
FRONT_IN_NEG = 8 Front Panel
Neg Edge
The trigger is asserted with a
negative edge polarity.
VXI0 = 0
VXI1 = 1
VXI2 = 2
VXI3 = 3
VXI4 = 4
VXI5 = 5
VXI6 = 6
VXI7 = 7
VXI Backplane
TTL Trigger
From any one of the
eight-shared VXI backplane
TTL triggers (TTLTRG0 -
TTLTRG7). The VXI backplane
TTL triggers follow the
Synchronous Trigger Protocol
as outlined in the VXIbus
Specification. This means that
the VXI backplane trigger line
must be asserted low for at
least 250 ns.
IN_MANUAL = 9 Manual From a software Trig In trigger
using
HPE6432_GenerateManualT
riggerInput() or from the
List Dialog Box using the
manual Trigger button.
IN_OFF = 15 Off Trig In triggers are disabled.