User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 3-43
Soft Front Panel Help
Configuration Dialog Box
Lock RF Attenuator
Lock RF Attenuator, available from the Configuration Dialog Box,
allows the RF attenuator (Option 1E1 Only) to be locked at its current
setting.
Related Topics
Soft Front Panel Help
Configuration Dialog Box Controls
Atten Lock Indicator
10 MHz Ref
10 MHz Ref, available from the Configuration Dialog Box, allows
selection of the internal 10 MHz reference or an external 10 MHz
reference.
If internal is selected, a 10 MHz reference signal is provided internally
by the synthesizer. If external is selected instead, a 10 MHz reference
signal greater than 0 dBm must be supplied through the 10 MHz In
connector that is available from the Hardware Front Panel.
Factory Preset Value: Internal
Related Topics
Soft Front Panel Help
Configuration Dialog Box Controls
10 MHz In
HPE6432_SetRefSource
Settling Time
Settling time is the period of time the synthesizer waits, following the
switch/blanking time, before producing a Trig Out trigger or a Sync Out
trigger. Settling time is used every time a new frequency or power is set
up in the synthesizer.