User`s guide

Agilent E6432A Microwave Synthesizer Users Guide 3-31
Soft Front Panel Help
ALC System Diagram
Uncoupled Operation
During uncoupled operation, the ALC and attenuation values are
changed independently.
In some applications, it is advantageous to control the ALC level and
step attenuator separately, using combinations of settings that are not
available in coupled operation. Uncoupled operation is accomplished as
follows:
by adjusting the ALC value from the main Soft Front Panel, only the
ALC level is changed
by adjusting the attenuation value from the main Soft Front Panel,
only the attenuation is changed
NOTE Uncoupled operation is always recommended when in external leveling
mode and must be used when specifying points in a list.
Related Topics
Soft Front Panel Help
Leveling (ALC) Controls
Understanding the ALC System