Service manual
E8408A Product Specifications 61
Appendix A
Safety/Environmental Specifications
This section describes safety and environmental specifications for the
E8408A mainframe, including:
• Environmental Specifications
• Electromagnetic Compliance/Safety Specifications
• Power Supply Protection
• Repair
Environmental
Specifications
Area Specification
Operating
Location
Indoor: Sheltered location where air temperature and humidity is controlled within this
product’s specifications and the equipment is protected against direct exposure to
climatic conditions such as direct sunlight, wind, rain, snow, sleet and icing, water
spray or splash, hoarfrost, or dew (typically indoors). Pollution degree 2.
Temperature Operating Temperature Range: 0°C to +55°C
Storage Temperature Range: -40 °C to +70° C
Humidity
(Non-condensing)
Operating Humidity Range:
- Up to 95% RH from 0°C to +40 °C
- Up to 65% RH from +40°C to +55°C
Storage Humidity Range:
- Up to 95% RH from 0°C to +55°C
- Up to 90% RH from +55 °C to +65°C
Shock End Use Handling:
Half sine waveform, <3 msec duration,
∆ v = 160 cm/sec minimum
Transportation:
Trapezoidal waveform,
∆ v = 605 cm/sec, 30 g minimum
Vibration Operating and Functional:
5 to 500 Hz, 0.0001 g
2
/Hz Spectral Density
Survival, Swept Sine:
5 to 500 Hz Resonance Search, 5 minute dwell on resonances at 0.5g
Survival, Random:
0.015 g
2
/Hz Spectral Density
Altitude Up to 3,000 m