Service manual

E8408A Product Specifications 59
Appendix A
Output Voltage
Power Input
Voltage Allowed Variation* Ripple/Noise DC Load* Induced Ripple/Noise*
+5V +0.25V / –0.125V 50mV 50mV
+12V +0.60V / –0.36V 50mV 50mV
-12V –0.60V / +0.36V 50mV 50mV
+24V +1.20V / –0.72V 150mV 150mV
-24V –1.20V / +0.72V 150mV 150mV
-5.2V –0.26V / +0.156V 50mV 50mV
-2V –0.10V / +0.10V 50mV 50mV
* Specifications apply at the backplane, 0 - 55
o
C.
Area Specification Comments
Input Voltage/
Frequency
90 Vac min - 140 Vac max
(47 Hz min - 440 Hz max)
90 Vac min - 264 Vac max
(47 Hz min - 66 Hz max)
Input Voltage 90 Vac to 264 Vac CAT II
280 VA max
Automatic Ranging
Inrush Current < 75A @ 264 Vac
< 35A @ 132 Vac
Cold start
Power Switch On/Standby
Chassis Ground Tap M4 x 0.7 threaded nut insert On rear of mainframe