Service manual

46 Servicing the Mainframe Chapter 3
Problem Isolation
Table 3-2 lists primary mainframe symptoms, possible cause(s), and
suggested actions. Table 3-1 shows replacement part numbers.
Replacing Mainframe Components
This section gives guidelines to replace E8408A mainframe components,
including:
Removing Mainframe Accessory Items
Removing the Mainframe Cover
Replacing the Power Supply Assembly
Replacing the Monitor Board
Replacing the Fan Assembly
Replacing the Front Panel Tapped Strip
Replacement Power Cords
Table 3-2. Isolating Problems in the E8408A Mainframe
Symptom Possible Cause Action
No mainframe
power
Power Supply fuse blown
or defective Power Supply
- There are NO replaceable power line fuses on mainframe.
- If the mainframe power supply fails to operate, replace the
Power Supply Assembly.
Short between some
supplies can latch the
Power Supply OFF
- Switch to Standby mode, and then remove AC power.
- Recovery can occur when AC power is reapplied and the
mainframe Power switch is switched to ON.
- If problem remains, replace Power Supply Assembly.
Flashing Voltages
Indicator
Backplane voltage(s) may
be incorrect
- Cycle mainframe Power switch.
- Remove VXI instruments to ensure they are not the cause.
- Check cable connections at Monitor and Backplane Board.
- Replace Power Supply Assembly.
Flashing Temp
Indicator
Power supply temperature
is high.
- Mainframe is nearing automatic shutdown.
- Check for proper air flow and clearance around the fan.
- Check power requirements of installed VXI instruments.
- Replace Fan Assembly as required.
- Replace Power Supply Assembly as required.
Flashing Fan
Indicator
Fan has failed or fan
connection loose
- Turn off mainframe to avoid possible overheating.
- Check fan connections to P3 on Backplane Board.
- Replace Fan Assembly as required.
Monitor indicators
and fan remain
OFF when Power
switch is pressed.
Defective Monitor Board
(E8408-66502) or loose
cable to Monitor Board
- Check cable connections at Monitor and Backplane Board.
- Replace Monitor Board as required.
Slot 0 Card
SYSFAIL
Indicator
VXI Instrument failure - The Slot 0 card (E1406A, E8491A, etc.) SYSFAIL
indicator will turn ON and remain ON when a VXI
instrument installed in the mainframe fails.