User`s manual
DIAGnostic
The DIAGnostic subsystem contains commands used to check the module
configuration, enable or disable all outputs simultaneously, change a channel’s
mode without opening and closing the output relay, and check the state of the trigger
on all channels.
Subsystem Syntax DIAGnostic
:CALibration
:OUTPut
[:MODE] <
path>
[:MODE]?
:CONFigure?
:OUTPut
:ALL
[:STATe] <
state>
:SOURce
n
:FUNCtion
:MODE <
mode>
:TRIGgered?
DIAGnostic:CALibration:OUTPut[:MODE]
DIAGnostic:CALibration:OUTPut[:MODE] <path> sets the voltage calibration
path for the module to either the CAL output terminals or at each output channel.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
path> discrete CALBus or CHANnel none
Comments • The typical calibration mode for the module is through the CAL output terminals.
Following
*RST or power-on, the default calibration mode is through the CAL
output terminals (and this command is not needed).
• This command with the CHANnel parameter, sets calibration to be performed at
each channel’s output terminals (and so includes the channel output relay
contacts).
• This command must be sent before any CALibrationn:CONFigure:VOLTage? or
CALibrationn:CONFigure:CURRent? query.
• *RST State: Calibration paths set to use the CAL output terminals.
DIAGnostic
76 Agilent E1418A SCPI Command Reference Chapter 3