User`s manual

DIAGnostic
The DIAGnostic subsystem contains commands used to check the module
configuration, enable or disable all outputs simultaneously, change a channel’s
mode without opening and closing the output relay, and check the state of the trigger
on all channels.
Subsystem Syntax DIAGnostic
:CALibration
:OUTPut
[:MODE] <
path>
[:MODE]?
:CONFigure?
:OUTPut
:ALL
[:STATe] <
state>
:SOURce
n
:FUNCtion
:MODE <
mode>
:TRIGgered?
DIAGnostic:CALibration:OUTPut[:MODE]
DIAGnostic:CALibration:OUTPut[:MODE] <path> sets the voltage calibration
path for the module to either the CAL output terminals or at each output channel.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
path> discrete CALBus or CHANnel none
Comments The typical calibration mode for the module is through the CAL output terminals.
Following
*RST or power-on, the default calibration mode is through the CAL
output terminals (and this command is not needed).
This command with the CHANnel parameter, sets calibration to be performed at
each channel’s output terminals (and so includes the channel output relay
contacts).
This command must be sent before any CALibrationn:CONFigure:VOLTage? or
CALibrationn:CONFigure:CURRent? query.
*RST State: Calibration paths set to use the CAL output terminals.
DIAGnostic
76 Agilent E1418A SCPI Command Reference Chapter 3