User`s manual
Setting the Logical Address Switch
Figure 1-4 shows the logical address switch location and setting. The logical
address switch factory setting is 72. Valid addresses range from 1 to 255.
When used with an Agilent Command Module the factory set address of
decimal 72 results in a module address of 9 (72 divided by 8). The module
address is used as a secondary address when using GPIB and a Command
Module. For example, the module address may be 7 09 09, where 7 is the
GPIB interface address, 09 is the command module address, and 09 is the
Agilent E1418A address.
The Agilent E1418A supports dynamic address configuration. When you
set the logical address to 255, the resource manager sets the logical address
programmatically.
Note When using an Agilent C-size mainframe with the Agilent E1405/E1406
Command Modules, the logical address setting must be a multiple of 8.
Figure 1-4. Setting the Logical Address
Chapter 1 Module Setup and Installation 19