User manual
92 Relay Multiplexer Command Reference Chapter 3
TRIGger:SLOPe
TRIGger:SLOPe <slope> is used to select the polarity of the output trigger.
For the E1460A, this command is not used.
Parameters
Comments Command Not Supported. Attempting to change the TRIGger:SLOPe to
anything other than NEG will generate an error.
TRIGger:SLOPe?
TRIGger:SLOPe? is used to query the polarity of the output trigger. For the
E1460A, this query always returns NEG.
Example Query Trigger Slope
TRIG:SLOP?
!Always returns NEG
TRIGger:SOURce
TRIGger:SOURce <source> specifies the trigger source to advance the
channel list during scanning.
Parameters
Comments Enabling the Trigger Source: The TRIGger:SOURce command only
selects the trigger source. The INITiate[:IMMediate] command enables
the trigger source.
Name Type Range of Values Default
<slope> discrete NEG NEG
Name Type Range of Values Default
BUS discrete *TRG or GET command IMM
ECLTrgn numeric ECL Trigger bus line 0 or 1 IMM
EXTernal discrete "Trig In" port IMM
HOLD discrete Hold Triggering IMM
IMMediate discrete Immediate Triggering IMM
Trgn numeric TTL Trigger bus line 0 - 7 IMM