User manual

Relay Multiplexer Command Reference 81Chapter 3
Comments Order of Command Execution: The [ROUTe:]SCAN:MODE and
[ROUTe:]FUNCtion commands must be executed before the
[ROUTe:]SCAN <channel_list> command.
[ROUTe:]SCAN:MODE versus [ROUTe:]FUNCtion:FRES: Measurement
mode is not supported when FUNCtion is set to WIRE1 (one-wire mode).
NONE and VOLT Mode: When selected, channel_list is setup for volts
measurements. VOLT mode is also used when making two-wire ohms
measurements using two-wire multimeters.
RES Mode: When selected, channel_list is setup for two-wire ohms
measurements. Control relay 0994 is closed when SCAN:PORT ABUS
is selected. When selected, the multimeter SENSE/SOURCE leads are
used to make the measurement. When using the HI/LO leads on a
multimeter to make the measurement, use the VOLT mode.
FRES Mode: When selected, channel_list is setup for four-wire ohms
measurements. When using four-wire ohms measurement mode, only
the lower bank (0-3) is specified with the [ROUTe:]SCAN <channel_list>
command. The upper bank pair (4-7) will automatically select the
specified channel. Selecting an upper bank (4-7) channel causes an
error.
*RST Condition: [ROUTe:]SCAN:MODE NONE
Example Selecting Four-Wire Ohms Measurements
This example selects four-wire ohms measurement mode (FRES) on
card #1 of a single-module switchbox.
FUNC 1,WIRE4
!Set mode to four-wire
TRIG:SOUR EXT
!Selects external trigger source
SCAN:MODE FRES
!Selects four-wire W scan mode
SCAN (@130:137)
!Scan channels 0 to 7 in bank 3 (paired
with channels 0 to 7 in bank 7)
INIT
!Starts scanning cycle