Relay Multiplexer Command Reference 63Chapter 3
ARM
The ARM subsystem selects the number of scanning cycles (1 to 32767)
for each INITiate command.
Subsystem Syntax ARM
:COUNt <number> MIN I MAX
:COUNt? [MIN | MAX]
ARM:COUNt
ARM:COUNt <number> MIN | MAX allows scanning cycles to occur a
multiple of times (1 to 32,767) with one INITiate command when
INITiate:CONTinuous OFF | 0 is set. MIN sets 1 cycle and MAX sets
32,767 cycles.
Parameters
Comments Number of Scans: Use only numeric values between 1 and 32767, MIN,
or MAX for the number of scanning cycles.
Related Commands: ABORt, INITiate:IMMediate
*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
This example sets a multiplexer module for 10 scans of channels 0
through 7 in bank 1 in a single-module switchbox.
ARM:COUN 10
!10 scans per INIT command
SCAN (@110:117)
!Scan channels 0 to 7 in bank 1
INIT
!Start scan, close channel 0
Name Type Range of Values Default
<number> numeric 1 thru 32,767 I MIN I MAX 1