User manual

Table 4-1. Agilent E1460A Common Problems
Problem
Type Symptom
Possible
Solutions
Self-test Errors Non-zero error code in response
to the *TST? command.
See Table 2-1 for information on
self-test errors.
Operator Errors Non-zero error code in response
to the SYST:ERR? command.
See Appendix C of the Agilent
E1460A User’s Manual for
multiplexer errors and causes.
See Appendix B of the
Agilent E1405 User’s Manual or
Agilent E1406 User’s Manual for
additional information on
operator errors.
Catastrophic
Failures
Not responding to commands.
Check logical address setting.
Check GPIB cables and
connections.
See “Testing the Assemblyin
this chapter.
Performance
Out of
Specification
Failing Closed-channel
Resistance Test (see Test 2-1 in
Chapter 2).
Check user wiring and test
connections.
Replace relays that correspond
to the channels that are failing
(see Table 4-3).
If most of the channels are near
or above the test limit (3.5 ),
replace the entire printed circuit
board (Agilent part number
E1460-66502).
Failing DC Isolation Test
(see Test 2-3 in Chapter 2).
Check user wiring and test
connections.
Remove dust from relay module
and terminal module printed
circuit boards.
50 Service Agilent E1460A Service Manual