User manual
High & Low
to Chassis
4. Hardware Connections
Note
Use any convenient chassis connection. The illustration shows the DMM
LO terminal connected to the outside of the Trig Out BNC on the E1406
Command Module.
5. Equipment Setup
• Send CLOS (@ss00:ss77) to the multiplexer to close all channel
relays.
6. DC Isolation Reading
• Trigger the DMM and record the reading.
Figure 2-6. High & Low to Chassis Isolation
Agilent E1460A Service Manual Verification Tests 29