User manual

Incoming
Inspection
Use the following steps as guidelines to perform initial (incoming)
inspection of the Agilent E1460A.
WARNING
To avoid possible hazardous electrical shock, do not perform
electrical tests if there are signs of shipping damage to the
shipping container or to the instrument.
1. If the Agilent E1460A is damaged, contact Agilent Technologies and
contact the carrier.
2. Install the Agilent E1460A in a VXI mainframe. Refer to the Agilent
E1460A Users’s Manual and the Agilent Series C Installation and
Getting Started Guide for additional information.
3. Perform the Functional Verification test and (optionally) the
Performance Verification tests. Refer to Chapter 2 in this manual.
4. If any of the tests do not pass, return the module to Agilent
Technologies. See Shipping Guidelines later n this chapter.
5. If all verification tests pass, the module is ready to use.
Agilent E1460A Service Manual General Information 13