Resistance
Adjustments
This procedure adjusts the E1411 multimeter 4-wire resistance measurement
accuracy.
NOTE
The DC Voltage adjustment and the AC Voltage adjustment MUST be
performed before the Resistance adjustment.
Equipment
Setup
1. Connect the equipment as shown in Figure 5-3
2. Set the E1411 as follows:
Reset Multimeter .................................................... *RST
Auto Zero ................................................................... ON
Power Line Cycles (PLC) .............................................. 1
Line Freq Reference (CAL:LFR) .......... 50 Hz or 60 Hz
Figure 5-3. Resistance Adjustments Setup
42 Adjustments Agilent E1411A/E1411B Service Manual