Service manual
AC Voltage
Adjustments
This procedure adjusts the E1411 AC voltage measurement accuracy.
NOTE
The DC Voltage adjustment MUST be performed before the AC Voltage
adjustment.
Equipment
Setup
1. Connect the equipment as shown in Figure 5-2
2. Set the E1411 as follows:
• Reset Multimeter ....................................................*RST
• Auto Zero .................................................................... ON
• Power Line Cycles (PLC) .............................................. 1
• Line Freq Reference (CAL:LFR ) ...... 50 Hz or 60 Hz
Figure 5-2. AC Voltage Adjustment Setup
Agilent E1411A/E1411B Service Manual Adjustments 39