Service manual
Table 4-1. Performance Test Record for the E1411 Multimeter (Page 3 of 3)
Model ________________________________ Report No. _______________________________ Date_________________
90 day Specifications
Test
No.
Test
Input
DMM
Range
Low
Limit
Measured
Reading
High
Limit
Meas
Uncert*
Test Acc
Ratio
(TAR)**
DC Voltage (Zero Volts Input) (Values in V)
4-1
0.0
0.0
0.0
0.0
0.0
0.113
0.910
7.270
58.100
300.000
-.000005
-.000015
-.000050
-.001000
-.005000
__________
__________
__________
__________
__________
+.000005
+.000015
+.000050
+.001000
+.005000
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
DC Voltage (DCV Input) (Values in VDC)
4-2
0.1
0.9
7.0
58.0
300.0
0.113
0.910
7.270
58.100
300.000
.099972
.899868
6.999250
57.990300
299.950000
__________
__________
__________
__________
__________
.100028
.900132
7.000750
58.009799
300.050000
.0000012
.0000044
.0000210
.0002820
.0017750
>10:1
>10:1
>10:1
>10:1
>10:1
AC Voltage (20 Hz, 60 Hz, 10 kHz, 5 kHz) (Values in VAC)
4-3
0.07
0.07
0.07
0.07
300.00
0.0875
0.0875
0.0875
0.0875
300.00
.067480
.069327
.069327
.067578
296.125000
__________
__________
__________
__________
__________
.072523
.070673
.070673
.072423
303.875000
.000020
.000016
.000016
.000016
.047000
>10:1
>10:1
>10:1
>10:1
>10:1
4-Wire Resistance (Values in Ohms)
4-4
1000
100000
1000000
2000
131000
1000000
999.730
99974.000
999740.000
__________
__________
__________
1000.270
100026.000
1000260.000
0.008
0.900
22.000
>10:1
>10:1
>10:1
* Measurement Uncertainty of Datron 4708 source for 90 days since calibration and 23
o
C ±1
o
C.
** TAR = multimeter accuracy/measurement uncertainty, rounded to nearest integer for <10:1 and >10:1 otherwise.
Agilent E1411A/E1411B Service Manual Verification Tests 33