Service manual
2. Set the E1411 multimeter as follows:
• Reset Multimeter ..................................................*RST
• Auto Zero ................................................................ON
• Power Line Cycles (PLC) ...........................................1
• Line Freq Reference (CAL:LFR) .......... 50Hz or 60Hz
NOTE
*RST sets Auto Zero to ON and Power Line Cycles to 1.
Test Procedure 1. Set the AC Standard Output to 0.07 Vac at 20 Hz sine wave.
2. Set the E1411 range to 79.5 mV (87.5 mV with 10% overrange)
using CONF:VOLT:AC .07.
3. Measure the AC input voltage with READ? and verify the
results are within specified limits (at the range selected for 1 PLC).
Record the results on the Performance Test Record.
4. Repeat steps 1 through 3 using the following AC Standard
voltage and frequency settings, and E1411 ranges:
Figure 4-3. AC Voltage Setup
Agilent E1411A/E1411B Service Manual Verification Tests 25