Service manual

Test 4-2: DC Voltage Test (DCV Input)
This test verifies DC Voltage accuracy on all five ranges with DC voltage
inputs.
Equipment Setup 1. Connect the equipment as shown in Figure 4-2.
WARNING
The DC Standard (Datron 4708, Option 10) can produce
dangerous voltages which are present on the terminals. Do not
touch the front (or rear) panel terminals unless you are sure no
dangerous voltage is present.
2. Set the E1411 multimeter as follows:
Reset Multimeter .................................................*RST
Auto Zero ................................................................ON
Power Line Cycles (PLC) ...........................................1
Line Freq Reference (CAL:LFR) ..... .....50Hz or 60Hz
Figure 4-2. DC Voltage (DCV Input) Setup
22 Verification Tests Agilent E1411A/E1411B Service Manual