Service manual
Multimeter
Serial Numbers
Multimeters covered by this manual are identified by a serial number prefix
listed on the title page. Agilent Technologies uses a two part serial number
in the form 0000A00000, where 0000 is the serial prefix, A is the country of
origin (A=USA) and 00000 is the serial suffix. The serial number prefix
identifies a series of identical instruments. The serial number suffix is
assigned sequentially to each instrument.
If the serial number prefix of your instrument is greater than the one listed
on the title page, a Manual Update (as required) will explain how to adapt
this manual to your instrument. If the serial number prefix of your
instrument is lower than the one listed on the title page, information
contained in Chapter 7- Manual Changes will explain how to adapt this
manual to your instrument.
Multimeter
Options
There are no electrical or mechanical options available for the E1411
multimeters. However, for the E1411B, you can order Option 1BN that
provides a MIL-STD-45662A Calibration Certificate. Contact your nearest
Agilent Technologies Sales and Support Office for information on Option
1BN.
Recommended
Test Equipment
Table 1-1 lists the test equipment recommended for testing, adjusting and
servicing the multimeter. Essential requirements for each piece of test
equipment are described in the Requirements column.
Table 1-1. Recommended Test Equipment
Instrument Requirements Recommended
Model
Use*
Controller, GP-IB GP-IB compatibility as defined by IEEE
Standard 488-1987 and the identical
ANSI Standard MC1.1: SH1, AH1, T2,
TE0, L2, LE0, SR0, RL0, PP0, DC0,
DT0, and C1, 2, 3, 4, 5
HP 9000 Series 300
or
IBM Compatible PC
with HP BASIC
A,F,
P,T
Mainframe Compatible with multimeter E1400B/T or E1421A/B A,F,
P,T
AC Standard Voltage Range 0.1 V to 300 V Datron 4708 with
Option 20
A,P
DC Standard Voltage Range 0.07 V to 300 V Datron 4708 with
Option 10
A,P
Resistance Standard
Values 1 k
Ω to 1 MΩ
Datron 4708 with
Option 30
A,P
Digital Multimeter General Purpose Voltage and Resistance 3458A T
*A = Adjustments, F = Functional Verification, P = Performance Verification Tests, T = Troubleshooting
12 General Information Agilent E1411A/E1411B Service Manual