Specifications

OUTPut
The OUTPut subsystem enables or disables the "Trig Out" port of the HP
E1300/E1301 Mainframe.
Subsystem Syntax OUTPut
[:STATe] <
state>
[:STATe]?
[:STATe] OUTPut[:STATe] <state> enables/disables the "Trig Out" port on the rear
panel of the HP E1300/E1301 Mainframe.
OUTPut[:STATe] ON | 1 enables
the port and
OUTPut[:STATe] OFF | 0 disables the port.
Parameters
Parameter Name Parameter Type Range of Values
state
Boolean 0 | 1 | ON | OFF
Comments Downloaded FET Multiplexer Scans: The OUTPut subsystem
does not work with downloaded FET multiplexer scan lists.
Downloaded scan lists are for use with the digital bus.
OUTPut
works with all other scanning modes.
Enabling "Trig Out" Port: When enabled, the "Trig Out" BNC is
pulsed each time a channel is closed during scanning.
Output Pulse: The pulse is a +5 V negative-going pulse.
"Trig Out" Port Reserved by Switchboxes: The "Trig Out"
resource is a system resource, and can only be reserved by one
switchbox instrument at a time.
Related Commands: [ROUTe:]SCAN, TRIGger:SOURce
*RST Condition: OUTPut[:STATe] OFF (port disabled)
Example Enabling "Trig Out" Port
OUTP ON Enables "Trig Out" port for pulse
output.
[:STATe]? OUTPut:[STATe]? queries the present state of the "Trig Out" port. The
command returns 1 if the port is enabled, or 0 if disabled.
Example Query "Trig Out" Port State
OUTP ON Enables "Trig Out" port for pulse
output.
OUTP:STAT? Query port enable state.
Chapter 5 Strain Gage Multiplexer Command Reference 79
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com