Specifications
:CONTinuous? INITiate:CONTinuous? queries the scanning state. With continuous
scanning enabled, the command returns 1. With continuous scanning
disabled, the command returns 0.
Example Query Continuous Scanning State
INIT:CONT ON Enables continuous scanning.
INIT:CONT? Query continuous scanning state.
[:IMMediate] INITiate[:IMMediate] starts the scanning cycle and closes the first channel
in the channel list. Successive triggers from the source specified by the
TRIGger:SOURce command advance the scan through the channel list.
Comments • Starting the Scanning Cycle: The INITiate[:IMMediate] command
starts scanning by closing the first channel in the channel list. A
trigger advances the scan through the channel list. An invalid
channel list generates an error (see the
[ROUTe:]SCAN command).
• Stopping Scanning Cycles: See the ABORt command.
Example Enabling a Single Scan
SCAN (@100:107) Sets channel list.
INIT Starts scanning cycle.
76 Strain Gage Multiplexer Command Reference Chapter 5