Specifications

Comments The strain gage wiring diagram used for this example is on page 28.
The diagram showing how the HP 3458A Multimeter is connected to
the FET strain gage multiplexer is on page 32.
When making strain measurements with an external voltmeter, the
following voltages are measured in the sequence shown:
bridge output voltage (Vout: channels 100 - 102) unstrained
bridge excitation voltage (Vs: channel 115) unstrained
bridge output voltage (Vout: channels 100 - 102) strained
bridge excitation voltage (Vs: channel 115) strained
These measurements are used to compute Vr, which in equation
form is defined as:
Vr = [(Vout/Vs)strained - (Vout/Vs)unstrained]
Vr is then substituted into the equation (Table 4-1 on page 52) for
1
4
bridge measurements.
The channel closed/measurement handshake sequence between the
FET multiplexer switchbox and the voltmeter occurs over the
multiplexer’s digital bus. When connected as shown in Figure 2-8,
the digital bus cable (HP Part Number E1411-80001) connects the
multiplexer’s "channel advance" and "channel closed" pins to the
voltmeter’s "VM Compl" and "Ext Trig" ports, respectively. Thus,
when a multiplexer channel is closed, a "channel closed" pulse from
the multiplexer is applied to the voltmeter’s "Ext Trig" port. When
the voltmeter measurement is complete, the voltmeter complete
signal triggers the multiplexer via the "channel advance" line.
Continued on Next Page
44 Making Strain Gage Measurements Chapter 3