Specifications

Single-Channel Bending Full Bridge Measurements
This program makes a bending full bridge strain measurement on channel 7
of multiplexer card number 1.
!Clear and reset the multimeter.
CLEAR 70903
OUTPUT 70903;"*RST"
!Specify the gage factor for the channel measured.
OUTPUT 70903;"STR:GFAC 2.115E-6,(@107)"
!Measure the channel’s unstrained reference.
OUTPUT 70903;"CAL:STR (@107)"
DISP "Add tension to the specimen; then press ’Continue’"
PAUSE
DISP ""
!Measure the strain on the channel. Enter and display the reading.
OUTPUT 70903;"MEAS:STR:FBEN? (@107)"
ENTER 70903;Strn_rdg
PRINT Strn_rdg
END
Comments The wiring diagram used for this example (FBENding) is on page 30.
When making full bridge measurements, the channels bridge
selection jumper must be in the "FULL" position.
Specifying the gage factor as 2.115E-6 returns the strain
measurements in microstrain.
A gage factor must be specified and an unstrained reference must be
measured for each channel.
40 Making Strain Gage Measurements Chapter 3