Specifications
Single-Channel 1/4 Bridge Measurements
This program makes a
1
⁄
4
bridge strain measurement on channel 0 of
multiplexer card number 1.
!Clear and reset the multimeter.
CLEAR 70903
OUTPUT 70903;"*RST"
!Specify the gage factor for the channel measured.
OUTPUT 70903;"STR:GFAC 2.11E-6,(@100)"
!Measure the channel’s unstrained reference.
OUTPUT 70903;"CAL:STR (@100)"
DISP "Add tension to the specimen; then press ’Continue’"
PAUSE
DISP ""
!Measure the strain on the channel. Enter and display the reading.
OUTPUT 70903;"MEAS:STR:QUAR? (@100)"
ENTER 70903;Strn_rdg
PRINT Strn_rdg
END
Comments • The wiring diagram used for this example is on page 28.
• Specifying the gage factor as 2.11E-6 returns the strain measurement
in microstrain.
• A gage factor must be specified and an unstrained reference must be
measured for each channel.
Chapter 3 Making Strain Gage Measurements 35