Specifications

Scan Control
Register
Reading the Scan Control Register allows you to check the scan control
settings.
base + 06
16
1514131211109876543210
Read F F 1 1 1 1 CNT
EN
IMM
EN
DBS
EN
CLR
SCN
CLR SCN Clear Scan List. A one (1) in this bit indicates the scan list is cleared. The
bit must be set back to zero (0) after it is cleared in order for another scan
list to be accepted.
DBS EN Digital Bus Enable. A one (1) indicates the Digital Bus trigger mode is
enabled. A zero (0) indicates it is disabled.
IMM EN Immediate Enable. A one (1) in this bit indicates the immediate trigger
mode is set.
CNT EN Continuous Enable. A one (1) in this bit indicates the continuous scanning
mode is set.
Scan Channel
Delay Register
Reading the Scan Channel Delay Register returns the SETTling:TIME delay.
base + 08
16
1514131211109876543210
Read F F 1111D3D2D1D0
D3 - D0
Data bits three to zero. The delay is 2
n
ยตsec and
n
has a range of 0 to 15
as determined by the settings of D3 through D0.
128 Strain Gage Register-Based Programming Appendix D