Guard Voltage
The Guard Voltage diagnostic accessed through internal channel 14
measures the guard voltage. This measurement verifies that the guard on
the relay strain gage multiplexers is driven to the necessary potential by the
Wagner Ground. The proper potential minimizes leakage currents caused
by low gage isolation resistance. Figure C-5 shows how the guard is driven
to a known voltage, and where that voltage is measured.
Specification The guard voltage should be
1
⁄
2
the bridge excitation voltage ± 2%.
Set Up To perform the guard voltage diagnostic, your system should be configured
as follows:
1. If the bridge excitation voltage is externally supplied, connect the
supply to the terminal module as shown in Figure C-1. Set the
voltage to 5.0V.
2. If desired, connect the bridge arrangement to the bridge completion
channels. Strain gages do not have to be connected for this
diagnostic.
CAUTION Use care when externally supplying the bridge excitation
voltage. Voltages greater than 5.4V may damage the bridge
completion resistors.
Figure C-5. Driving the Card