Specifications
Appendix C
Strain Gage Diagnostics
About This Appendix
This appendix covers the diagnostics associated with the strain gage
multiplexers. The diagnostics check configuration integrity and increase
measurement accuracy. The diagnostics in this appendix include:
• Leadwire Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 108
• Internal Half Bridge Voltage . . . . . . . . . . . . . . . . . . . . . . . . . Page 111
• Shunt Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 113
• Guard Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 116
• Bridge Excitation Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 118
Diagnostic
Channels
The multiplexer diagnostics are accessed through "internal" channels as
well as the "external" bridge completion channels. Table C-1 identifies the
internal channel and the diagnostic on that channel.
Table C-1. Strain Gage Multiplexer Diagnostic Channels
Channel Diagnostic
8/9 Leadwire Resistance. Channels 8 and 9 measure the voltage drop in the
leadwires from the strain gage to the bridge completion circuit, for bridge
completion channels 0 and 1. From this voltage and the value of the bridge
completion resistor, the leadwire resistance can be determined so that
leadwire desensitization corrections can be made.
10/11 Internal Half Bridge Voltage. Channels 10 and 11 measure the lower and
upper legs of the internal half bridge voltage. This checks the stability of the
1,000Ω bridge completion resistors.
12/13 Shunt Verification. Channel 12 places a resistor in parallel with the bridge
completion channel strain gage to simulate a known value of compressive
strain. Channel 13 places a resistor in parallel with the upper leg internal
bridge resistor to simulate a known value of tensile strain.
14 Guard Voltage. Channel 14 measures the guard voltage (Wagner Ground)
to determine if it is 1/2 of the bridge excitation voltage.
15 Bridge Excitation Voltage. Channel 15 measures the bridge excitation
voltage (Vs).
Appendix C Strain Gage Diagnostics 107