Service manual

HI to Chassis Leakage 1. Make Hardware Connections
Turn power supply and mainframe power OFF
Connect DMM, power supply, and resistor as shown in
Figure 2-8
Turn power supply and mainframe power ON
2. Check Direct Terminals Leakage
Send *RST to FET Multiplexer
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to Chassis, Direct
The DMM measurement should be less than 0.010 Vdc.
A measurement out of this range indicates a failure of the
FET Multiplexer
Agilent E1300B/E1301B
Figure 2-7. Negative HI to LO Leakage Connections
2-14 Verification Tests