Service manual

Repeat step 3 for channels 01 through 15
Use CLOS (@nncc), where nn is the card number and cc is the
channel number
LO Channel
Measurements 1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM as shown in Figure 2-3
Turn mainframe power ON
2. Check for Stuck Channels
Send *RST to the FET Multiplexer to open all channels
Trigger the DMM with TRIG SGL and note reading
Verify the reading is greater than 10 k
3. Check LO Closed Channel Resistance
Send CLOS (@nn00) to close channel 00, where nn is the card
number (typically 01)
Trigger the DMM with TRIG SGL and note the reading
Enter the result in Table 2-1 for channel 00 LO
Send OPEN (@nn00) to open channel 00, where nn is the card
number
4. Repeat for Channels 01 through 15
Repeat step 3 for channels 01 through 15
Use CLOS (@nncc) and OPEN (@nncc), where nn is the card
number and cc is the channel number
Agilent E1300B/E1301B
Figure 2-3. LO Channel Resistance Test Connections
Verification Tests 2-5