Service manual

re-wire each time the tests are performed. The Agilent E1351A terminal
block and case assembly part number is E1351-80001.
Test 2-1: Closed
Channel
Resistance Test
This test first verifies that no FET switches are stuck in the on condition and
then verifies that all channels meet the closed channel resistance
specification for the FET Multiplexer. The channel HI, channel LO, A Tree,
and B Tree switches are all independently tested.
HI Channel
Measurements
1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM as shown in Figure 2-2
Turn mainframe power ON
2. Check for Stuck Channels
Send *RST to the FET Multiplexer to open all channels
Trigger the DMM with TRIG SGL and note reading
Verify the reading is greater than 10 k
3. Check HI Closed Channel Resistance
Send CLOS (@nn00) to close channel 00, where nn is the card
number (typically 01)
Trigger the DMM with TRIG SGL and note the reading
Enter the result in Table 2-1 for channel 00 HI
Send OPEN (@nn00) to open channel 00, where nn is the card
number
4. Repeat for Channels 01 through 15
Agilent E1300B/E1301B
Figure 2-2. HI Channel Resistance Test Connections
2-4 Verification Tests