Service manual

Send CLOS(@nn15) to the FET Multiplexer to close channel
15 and tree B
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to Chassis, Tree B
Send *RST to the FET Multiplexer
5. Change Polarity
Turn power supply and mainframe power OFF
Connect DMM, power supply, and resistor as shown in Figure
2-9
Turn power supply and mainframe power ON
Set power supply output to +10 Vdc ± 0.1 Vdc
6. Repeat Steps 2 through 4
Record all results in Table 2-1 as
Negative Polarity, HI to Chassis
Agilent E1300B/E1301B
Figure 2-9. Negative HI to Chassis Leakage Connections
2-16 Verification Tests