User`s manual

OUTPut
The OUTPut subsystem enables or disables the "Trig Out" BNC port of the
Agilent E1300/E1301 Mainframe.
Subsystem Syntax OUTPut
[:STATe] ON |OFF |1 |0
:STATe?
[:STATe] OUTPut[:STATe] ON |OFF |1 |0 enables/disables the "Trig Out" BNC port
on the Agilent E1300/E1301 Mainframe rear panel. OUTPut[:STATe] ON
|1 enables the port and OUTPut[:STATe] OFF |0 disables the port.
Parameters
Parameter Name Parameter Type Range of Values
ON |OFF |1 |0 boolean 0 |1 |ON |OFF
Comments
Enabling "Trig Out" Port: When enabled, the "Trig Out" BNC is
pulsed each time a channel is closed during scanning. When
disabled, the "Trig Out" BNC is not pulsed.
Output Pulse: The pulse is a +5 V negative going pulse.
"Trig Out" Port Shared by Switchboxes: When enabled, the
"Trig Out" BNC may be pulsed by any switchbox each time a
channel is closed in a switchbox during scanning. To disable the
output for a specific switchbox, send the OUTPut[:STATe] OFF or
OUTPut[:STATe] 0 command for that switchbox.
Related Commands: [ROUTE:]SCAN, TRIGger:SOURce.
*RST Condition: OUTPut[:STATe] OFF (port disabled).
Example Enabling "Trig Out" BNC Port
OUTP ON Enables "Trig Out" BNC for pulse
output
:STATe? OUTPut:STATe? queries the present state of the "Trig Out" BNC port. The
command returns 1 if the port is enabled, or 0 if disabled.
Example Query "Trig Out" BNC Port State
OUTP ON Enables "Trig Out" BNC for pulse
output
OUTP:STAT? Query port enable state
48 Agilent E1346A Multiplexer Command Reference Chapter 5