ABORt
The ABORt subsystem stops a scan in progress when the scan is enabled via
the interface, and the trigger modes are TRIGger:SOURce BUS or
TRIGger:SOURce HOLD.
Subsystem Syntax ABORt
Comments
• ABORt Operation invalidates the current channel list and sets
ARM:COUNt 1 (one scanning cycles per INITiate command), sets
INITiate CONTinuous OFF (no continuous scanning cycles), and
sets TRIGger:SOURce IMMediate (continuous internal triggering).
• Stopping Scans Enabled from Interface: When a scan is enabled
from the interface, use an interface CLEAR command or the Agilent
E1301 front panel "Reset Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS or TRIGger:SOURce HOLD, use ABORt or
the Agilent E1301 front panel "Reset Instr "or "Clear Instr" keys to
stop the scan.
• Stopping Scans Enabled from Front Panel: When a scan is
enabled from the Agilent E1301 front panel execute *RST over the
interface or the front panel "Reset Instr" or "Clear Instr" keys to stop
the scan.
• Related Commands: ARM, INITiate:CONTinuous,
[ROUTe:]SCAN, TRIGger.
Example Stopping a Scan with ABORt
TRIG:SOUR BUS Bus is trigger source
INIT:CONT ON Set continuous scanning
SCAN (@100:147) Sets channel list
INIT Starts scanning cycle
.
ABOR Aborts scan in progress
42 Agilent E1346A Multiplexer Command Reference Chapter 5