Service manual

Table 2-2. Performance Test Record (Page 3 of 3)
Test No/Description Minimum*
Value
Measured Value (V) Maximum
Value **
Meas
Uncert
Test Acc
Ratio (TAR)
2-1. Closed Channel Resistance (Values in Ohms) (cont’d)
HI Path
Resistance
Channel 36 & 91
Channel 37 & 91
Channel 38 & 91
Channel 39 & 91
Channel 40 & 92
Channel 41 & 92
Channel 42 & 92
Channel 43 & 92
Channel 44 & 92
Channel 45 & 92
Channel 46 & 92
Channel 47 & 92
LO Path
Resistance
Channel 90
Channel 91
Channel 92
G Path Resistance
Channel 90
Channel 91
Channel 92
Channel 93 Path
Resistance
Channel 93 HI
Channel 93 LO
Channel 93 G
___________________
___________________
___________________
___________________
___________________
___________________
___________________
___________________
___________________
___________________
___________________
___________________
___________________
____________________
____________________
____________________
____________________
____________________
____________________
____________________
____________________
____________________
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
2.0
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
8.00E-5
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
2-2. DC Isolation (Values in Ohms) ***
HI to CHASSIS
HI to LO
HI to G
1E9
1E9
1E9
___________________
___________________
___________________
6.0E6
6.0E6
6.0E6
NA
NA
NA
*Single-sided specification - Minimum value does not apply.
**Value listed is for a single relay contact. All channels measured include two relay contacts. Refer to Chapter 4 for troubleshooting
information.
***Single-sided specification - Maximum value does not apply.
Verification Tests 2-23