User`s manual
Relay Multiplexer Command Reference 51
ARM
The ARM subsystem selects the number of scanning cycles (1 through 32767) for
each
INITiate command.
ARM
:COUNt <
number> MIN|MAX
:COUNt? [MIN |MAX]
:COUNt
ARM:COUNt <number> MIN|MAX allows scanning cycles to occur a multiple
of times (1 to 32767) with one
INITiate command and when
INITiate:CONTinuous OFF|0 is set.
Parameters
Comments • Number of Scans: Use only values between 1 to 32767 for the number of
scanning cycles (default is 1).
• Related Commands: ABORt, INITiate[:IMMediate]
• *RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
ARM:COUN 10
Sets 10 scanning cycles.
SCAN (@100:115)
Sets channel list.
IINIT
Starts scanning cycle.
Parameter Name Parameter Type Range of Values
<
number>
numeric 1–32767|MIN|MAX