User`s manual

50 Relay Multiplexer Command Reference
ABORt
The ABORt subsystem stops a scan in progress when the scan is enabled
via the interface, and the trigger modes are
TRIGger:SOURce BUS or
TRIGger:SOURce HOLD
.
Subsystem Syntax ABORt
Comments ABORt Operation invalidates the current channel list and sets ARM:COUNt 1
(one scanning cycles per
INITiate command), sets INITiate CONTinuous OFF
(no continuous scanning cycles), and sets
TRIGger:SOURce IMMediate
(continuous internal triggering).
Stopping Scans Enabled from Interface: When a scan is enabled from the
interface, use an interface
CLEAR command or the E1301 front panel "Reset
Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS or TRIGger:SOURce HOLD, use ABORt or the
E1301 front panel "Reset Instr" or "Clear Instr" keys to stop the scan.
Stopping Scans Enabled from Front Panel: When a scan is enabled from the
E1301 front panel, execute
*RST over the interface or the front panel "Reset
Instr" or "Clear Instr" keys to stop the scan.
Related Commands: ARM, INITiate:CONTinuous, [ROUTe:]SCAN, TRIGger
Example Stopping a Scan with ABORt
TRIG:SOUR BUS
Bus is trigger source.
INIT:CONT ON
Set continuous scanning.
SCAN (@100:115)
Sets channel list.
INIT
Starts scanning cycle.
ABOR
Aborts scan in progress.