22 Configuring the Relay Multiplexer Modules Chapter 2
Connecting the Analog Bus
Figure 2-2 shows how to connect the analog bus between multiple
multiplexer modules and to the E1326 Multimeter. Use the cables shipped
with the multiplexer modules to connect the analog bus of the multiplexer
modules. Use the cable shipped with the E1326 Multimeter to connect the
analog bus of the multiplexer module to the Multimeter input.
These cables are needed to connect the MUX when mounted in the E1403
Module Carrier to the E1411A/B DMM in a C-size cardcage. They must be
ordered separately.
1. Connecting E1411 to RELAY MUXs: use E1326-61611
(long analog bus cable).
2. Connecting RELAY to RELAY or RELAY to FET MUX:
use E1400-61605.
Figure 2-2. Analog Bus Connections Between Multiplexer/Multimeter Modules