Service manual
Recommended Test Equipment
Table 1-1 lists the test equipment recommended for testing, adjusting, and
servicing the relay multiplexers. Essential requirements for each piece of
test equipment are described in the Requirements column.
Table 1-1. Recommended Test Equipment
Instrument Requirements Recommended
Model
Use*
Controller, HP-IB HP-IB compatibility as defined by
IEEE Standard 488-1987 and the
identical ANSI Standard MC1.1:
SH1, AH1, T2, TE0, L2, LE0, SR0,
RL0, PP0, DC0, DT0, and C1, 2, 3, 4,
5.
HP 9000 Series 300
or
IBM compatible PC with
BASIC
F,O,
P,T
Mainframe Compatible with multiplexer E1300A, E1301A,
E1302A or E1401B/T,
E1421B (requires
E1406A)
F,O,
P,T
Digital Multimeter
2-wire ohms (up to 1 GΩ)
4-wire ohms
3458A or
34401A
O,P,T
* F = Functional Verification, O = Operation Verification Tests,
P = Performance Verification Tests, T = Troubleshooting
Inspection/Shipping
This section contains initial (incoming) inspection and shipping guidelines
for the multiplexers.
Initial Inspection Use the steps in Figure 1-2 as guidelines to perform initial inspection of a
relay multiplexer. Performance Verification tests are optional.
WARNING To avoid possible hazardous electrical shock, do not perform
electrical tests if there are signs of shipping damage to the
shipping container or to the instrument.
Chapter 1 General Information 17