User`s manual

Agilent E1330B Digital I/O Module Command Reference 75
Chapter 5
DIGital:CONTroln[:VALue]
[SOURce:]DIGital:CONTroln[:VALue] <value> sets or clears the control line on
the selected port n.
Parameters
Comments This command is used to create custom handshakes when the HANDshake is
set to
NONE.
:CONTroln is the keyword used for commands relating to the control (CTL)
line at port n. The port number n must be the last character of the keyword
without spaces.
The control line is used with the flag line to handshake data to and from
peripherals.
Related Commands: [SOURce:]DIGital:CONTroln:POLarity,
[SOURce:]DIGital:CONTroln:POLarity?, [SOURce:]DIGital:FLAGn:POLarity,
[SOURce:]DIGital:FLAGn:POLarity?
*RST Condition: Clears the control line; i.e., sets the control line to logical 0.
Example DIG:CONT2 1 sets the 8-bit port 2 control line true.
DIGital:CONTroln[:VALue]?
[SOURce:]DIGital:CONTroln[:VALue]? reads the state of the control line on port
n and returns a 0 or 1, indicating the logical condition of the CTL line.
Parameters
Comments This command is only available when using the downloaded SCPI driver.
This command is used to create custom handshakes when the HANDshake is
set to
NONE.
The condition of the CTL line returned by this command is the logical true
value set by the
DIGital:CONTroln:POLarity command.
Example DIG:CONT2? returns the current state of the 8-bit port 2 control line.
Parameter Name Parameter Type Range of Values Default
CONTrol
n
Numeric none, 0, 1, 2, or 3 0
<
value>
Boolean 0 or 1, OFF or ON None
Parameter Name Parameter Type Range of Values Default
CONTrol
n